Many clinicians are not aware that high serum levels of lactate dehydrogenase (LDH) can affect enzymatic assays for bicarbonate, resulting in falsely elevated reported bicarbonate levels. To evaluate this problem, researchers identified 549 patients for whom measured bicarbonate (by an enzymatic Beckman Coulter assay), calculated bicarbonate (from blood gas measurements), and LDH were all drawn within an hour of each other at a large medical center. For this observational study, researchers considered bicarbonate levels calculated from blood gases to be the reference values.
When LDH was <2000 U/l, the mean difference between measured and calculated bicarbonate was negligible. In contrast, the mean difference was 3.2 mmol/l…